COMPUTER INTERFACE FOR MULTIRAE PLUS
5 - 14
Back Light:
This option allows the user to choose to turn on and off the
LCD back light manually or automatically based on the
ambient light sensor input. If
Manual
mode is selected, the
user can also enter a time out period so that the back light
will be turned off automatically after the time out period. If
Auto
is selected, the ambient light sensor will automatically
turn on and off the backlight.
Clock from PC:
This option allows the PC clock to be downloaded to the
MultiRAE PLUS monitor so that the user does not need to
manually set the clock in the MultiRAE PLUS monitor.
Pump Speed:
This option allows the user to choose the high pump speed
or low pump speed for the MultiRAE PLUS monitor.
Measure:
The user can choose to record the average or peak (Max)
reading of all the sensors during each datalog period.
Datalog Interval:
The datalog interval is the rate at which datalog samples are
recorded while the datalog is running. The unit of datalog
interval is second. The datalog interval determines how often
a new sample is calculated and stored to the datalog. The unit
of a datalog interval is one second.