PROGRAMMING OF MULTIRAE
4 - 12
Note: the readings should be very close to the span gas
values. After about 5 seconds pause, the display will
show the message “Span Cal Done! Turn Off Gas”
3. This completes the multiple sensor calibration procedure
and moves to the next sub-menu item.
4. Turn off the flow of gas. Disconnect the calibration
adapter from the
MultiRAE PLUS
Monitor.
5. If the sensor failed the calibration, the sensor name and
an error message “failed, continue?” will appear. Press
the
[N/-]
or
[MODE]
key to abort the calibration and
move directly to the next sub-menu item. Press the
[Y/+]
key to continue the next sensor calibration. In
either case, the calibration data for the current sensor is
not changed.
6. From Step 1, if the
[N/-]
key is pressed, the display
shows all the sensor names which are selected for
multiple sensor calibration with the cursor blinking at the
first sensor location.
TOX1
VOC
TOX2
CO* - - - -
H
2
S*
LEL* OK?
O
2
LEL
OXY
Press the
[Y/+]
key to select the sensor and the
[N/-]
key
to de-select the sensor. A previously selected sensor
will show an “*” next to the sensor name. A previously
de-selected sensor will not show an “*” next to the
sensor name.