PROGRAMMING OF MULTIRAE
4 - 26
4.6.4 Select Data Type
The user can choose to store either the average or the peak
value during each datalog period.
1. “Select Data Type?” is the fourth sub-menu item in Table
4.5.
2. Press the
[Y/+]
key, the display shows the current data
type: “Data Type = Average?”
3. Press the
[Y/+]
key to accept the currently displayed data
type. Press
[N/-]
key to change to the other data. type:
“Data Type = Peak?” Press
[MODE]
key to exit this
sub-menu and move on to the next sub-menu.
4. If there is any change to the existing selection, pressing
the
[y/+]
key, the display will show “Save?” Then, press
the
[Y/+]
key to accept or the
[N/-]
key to discard and
move to the next sub-menu.
Note: It is necessary to select “Average” data type in order
to calculate the correct STEL and TWA data for
datalogging report (see Section 5.9).