OPERATION OF MULTIRAE PLUS
2 - 18
Alarm Signal Latching:
It is possible to setup the MultiRAE PLUS from a PC or in
programming mode so that when an alarm condition occurs,
the alarm signals stay on even after the alarm condition is no
longer present. This is called the “latching alarm” mode.
The alternative mode is to automatically reset the alarm
signal when the alarm condition is cleared. The default
alarm mode is automatic reset (see Chapter 4 or 5 for
details on how to set alarm mode).
Alarm Signal Summary:
The various alarm signals are summarized here:
Condition
Alarm Signal
Gas exceeds “High
Alarm” limit
3 beeps/flashes per second
plus sensor name on LCD
Gas exceeds “Low
Alarm” limit
2 beeps/flashes per second
plus sensor name on LCD
Gas exceeds STEL
1 beep/flash per second plus
sensor name on LCD
Gas exceeds TWA
1 beep/flash per second plus
sensor name on LCD
Negative Drift or
Over Range
3 beeps/flashes per second
plus NEG or OVR name on
LCD
Pump failure
3 beeps/flash per second plus
“Pump” message on LCD
LEL sensor off
3 beeps/flash per second plus
LEL sensor name and “Off”
message on LCD