TROUBLESHOOTING
8 - 2
8.1 Special Diagnostic Mode
To place the monitor into the special diagnostic mode, first
turn the monitor off. Next push and hold the
[Y/+]
key, while
still holding the
[Y/+]
key, also push and hold the
[MODE]
key for at least two seconds. After the required delay,
release both keys, the monitor will now go through the
normal start-up display sequence and display a “Diagnostic
mode” message. At the end of the special mode start-up, the
display will show 1 to 5 numbers with a message “Raw.”
These numbers are raw sensor readings without calibration.
Other critical parameters of the monitor are available as
shown in Table 8.1. They can be reached by pressing the
[MODE]
key repeatedly until the desired parameters are
displayed.