OPERATION OF MULTIRAE PLUS
2 - 21
2.7 Preset Alarm Limits and Calibration
The MultiRAE PLUS Multi-gas Monitor is factory
calibrated with standard calibration gas, and is programmed
with default alarm limits as listed below. (refer to Chapter 4
or Chapter 5 for programming procedures if new calibration
or alarm limits is required).
Table 2.2
Factory Calibration and Preset Alarm Limits
Gas
ppm
Cal Gas /
Balance
Unit
TWA
STEL Low High
CO
50 / Air
ppm
35
100
35
200
H
2
S
25 / N
2
ppm
10
15
10
20
SO
2
5 / N
2
ppm
2
5
2
10
NO
25 / N
2
ppm
25
25
25
50
NO
2
5 / Air
ppm
1
1
1
10
Cl
2
10 / N
2
ppm
0.5
1
0.5
5
O
2
20.9 / N
2
%Vol
-
-
19.5 23.5
CH
4
50 / Air
%LEL
-
-
10
20
HCN
10 / N
2
ppm
5
5
5
50
NH
3
50 / N
2
ppm
25
35
25
50
PH
3
5 / N
2
ppm
0.3
1
1
2
VOC*
100 / Air
ppm
10.0 25.0
50.0 100
*
Note
: 100 ppm isobutylene gas is used for VOC gas
calibration