PROGRAMMING OF MULTIRAE
4 - 8
Calibration Sub-Menu
Fresh Air Calibration?
Multiple Sensor Calibration?
Single Sensor Calibration?
Modify Span Gas Value?
Change LEL/VOC Span Gas?
The preferred calibration method for monitors with pumps is
to place the calibration gas in a gas sample bag (Tedlar
bag). Connect the bag to the MultiRAE PLUS gas inlet port
when you are ready to perform the calibration. Users may
find it more convenient to connect the MultiRAE PLUS
directly to the gas bottle using a flow-on-demand type
regulator. Calibration through such a regulator will produce
similar results to the sample bag method with most gases,
but flow-on-demand regulators are not recommended with
H
2
S. In an emergency, the MultiRAE PLUS can also be
calibrated directly from a fixed flow regulator with a flow
rate between 0.5 and 1.0 liters per minute.
MultiRAE PLUS diffusion monitors must be calibrated
using a fixed flow regulator with a flow rate between 0.5 and
1.0 liters per minute. Diffusion monitors are supplied with a
special calibration adapter that covers the gas diffusion port.
The calibration adapter links up the inlet port of the
MultiRAE PLUS Monitor to the gas sample bag (see Figure
4-1).