Saving and loading of pump configuration
Remote loading
Connect the hand held dislay module onto the pump to be configured.
Take the control of the pump from the hand-held module display (* at
cover left of the display).
Load the configuration previously saved in the hand held display
module into the pump by selecting the
CONFIG LOAD
menu from
MANAGEMENT
menu.
The former configuration is automatically reset.
All pump parameters are loaded remotely, except the following
parameters:
buzzer,
unit of temperature,
unit of pressure,
RS link,
time and date,
language.
Loading parameters may result in creating software conflicts.
After loading, configure manually the following parameters (temperature
unit, pressure unit, serial link, time and date).