GB 02878 - Edition 03 - Oct 13
3/3
B 60
Connection to the pumping circuit
At the pump inlet
To get the best pumping speed, ensure that the vacuum line is the
minimum length possible and has an internal diameter not less than the
inlet connection of the pump.
Connection
A103P
A603P / A1003P
Inlet connection
DN50 ISO-KF
DN100 ISO-K
Several fitting accessories are available in the adixen catalog.
Incorporate flexible stainless steel flanged bellows in the vacuum pipeline to
reduce the transmission of vibration.
Make sure that any vacuum accessories connected to the inlet of the pump
can withstand a negative pressure of 100 kPa relative to atmospheric
pressure.
adixen Vacuum Products - Operating instructions - A3P Series