D 30
Diagnosis and trouble shooting
GB 04155 - Edition 04 - Oct 13
INCIDENT
CAUSE
TROUBLESHOOTING
Warning
W01 EXHAUST PRESSURE
Pump does not start
Exhaust pressure > 1400 hPa
Check the exhaust pipe.
Other problem
Call the customer service center.
Pump is running
Exhaust pressure > 1250 hPa
Check the N
2
flowrate setting
value
B 40
.
Check the exhaust pipe.
Other problem
Call the customer service center.
Hazard
D01 EXHAUST PRESSURE
Pump does not start
Exhaust pressure > 1400 hPa
Check the exhaust pipe.
Other problem
Call the customer service center.
Pump is running
Exhaust pressure > 1990 hPa
Check the N
2
flowrate setting
value
B 40
.
Check the exhaust pipe.
Other problem
Call the customer service center.
Hazard
D42 SAFETY EXH PRESS
Exhaust pressure > 2000 hPa
Check the sensor connection
on the «Press switch» connector
D 10
.
Check the exhaust pipe.
Other problem
Call the customer service center.
Hazard
D07 FREQ. CONVERT. 50
Call the customer service center.
Display is showing a defect number