GB 03534 - Edition 01 - November 07
1/1
E 20
Increase of the maintenance parameters
DEFINITION MENU
SETTING
MAINTENANCE
MANAGEMENT
ENTER PASSWORD
« 0 «
OVERHAUL. WARNING
EXHAUST WARNING
PART. MAINT. WARNING
Press the
SET
key.
Press the key
+
and
-
until the code is obtained
(factory setting = 0)
Validate with the
ENTER
key.
Press the
+
key repeatedly to access to
MAINTENANCE
menu.
Validate with the
ENTER
key.
Press the
+
key repeatedly to access to the
choosen menu.
Validate with the
ENTER
key.
Increase of the
maintenance
parameters
PART. MAINT. WARNING
17000
MIN: 1000 H
MAX: 34000 H
The fresh screen appears.
Use the
+
keys to increase the maintenance
duration time.
Validate with the
ENTER
key.
Return to the main menu and exit the setting
mode using the
SET
key.
When the maintenance of an element has been done,
it is necessary
to increase to the next maintenance time to stop the warning
of this
element.
The zero-set-up of the maintenance parameters will be done in a service
center during a total maintenance.
Table of maintenance
17000
34000
0
Überholung
Zeit in Stunde
17000 h
oder 2 Jahren
reset to 0
17000 h
oder 2 Jahren
Zähler-Pumpe
(im Servicecenter)
Überholung
reset to 0
Überholung
(im Servicecenter)
Access to the parameter
programing
Access to choosen parameter
menu
adixen Vacuum Products - Operating instructions - A3P Series