GB 02907 - Edition 02 - November 07
1/1
E 14
Disconnecting the pump from the
installation
Turning off the pump
Turn off the pump, turn the main switch to «0».
Turn off the customer main circuit breaker and lock out tag out per
facilities procedures.
Disconnect the main wires from the electrical connector.
Description
A 20
Disconnecting the
facilities
Disconnect all electrical connectors.
Disconnect the water inlet and outlet quick connectors.
Disconnect the N
2
purge connector.
Disconnect the pump from the inlet piping.
Cap the equipment inlet piping.
Disconnect the secondary exhaust port from customer application.
Disconnect the pump from the exhaust piping.
Cap the equipment exhaust piping.
Description
A 20
Disconnect the pump
from the installation
Remove the anti-earthquake squares (if equipped).
Raise the four locking screw jacks.
Remove the pump from the installation.
Study the preliminary precautions
E 11
.
adixen Vacuum Products - Operating instructions - A3P Series