GB 02882 - Edition 03 - Januar
y 2010
1/1
B 100
EMO plug connection
(on A603P / A1003P pump models)
Presentation
The «EMO» plug allows the control or remoting of the emergency stop
function from the front panel of the unit.
Connector location
The emergency stop connection is done via the «EMO connector»,
located on the rear side of the pump.
Cover plug wiring
Contact
Function
Input
2-3
Contact opened, emergency stop command is activated.
Available output
1-4
This contact opens when emergency stop is activated.
Rear side
A603P / A1003P Pumps
EMO connector
Plug wiring
All dry contacts are rated 50V - 1A.
7IRINGVIEWEDUNDERSOLDEREDSIDE
Study the preliminary precautions
B 00
.
In local mode, the pump can run only if the cover plug (delivered with the
pump) is fitted on the «EMO connector».
When the emergency stop is interfaced, if the remote mode is not used, it is
necessary to connect the «remote cover plug».
When units containing control circuits are equipped with dry contacts
outputs, it is the customer’ responsibility to wire these outputs in compliance
with safety extra low voltage installation and security standards. It concerns
«Remote control», «emergency stop» and «RS232/485» connectors.
adixen Vacuum Products - Operating instructions - A3P Series