2/3
B 60
GB 02878 - Edition 03 - Oct 13
Connection to the pumping circuit
Secondary exhaust
enclosure
(Semi configuration)
The purpose of the secondary exhaust enclosure is to exhaust gas and
vapors that might leak from the pump.
A103P / A603P / A1003P
Flow
22 cfm
Differential pressure *
0.3 hPa / 0.125» wg
Exhaust
Ø 50 mm
* Measured at the secondary exhaust port
Connection procedure for
pump equipped in Semi
configuration
Remove the grey junck ring and the plate from the exhaust box by
unscrewing the 4 screws Chc M4x12.
Disconnect the 2 terminals of differential pressure sensor cable from the
differential pressure sensor.
Remove the differential pressure sensor from the
upper box by unscrewing the 2 screws M3.
Check the presence of the O-ring of the differential
pressure sensor.
Remove the upper box from the pump by unscrewing
the 4 screws Chc M4x12.
Place the plate then the grey junck ring on the
customer pipe.
Place the centering ring DN25 equipped with
O-ring on the exhaust port and connect the customer
pipe.
Secure it with the connect clamp.
Place the upper box.
Secure it with 4 screws Chc M4x12.
Replace the differential pressure sensor on the upper box and secure it
with the 2 screws M3.
Connect the 2 terminals of differential pressure sensor cables to the
differential pressure sensor.
Push the grey junck ring and the plate against the exhaust box.
Hold them in position with the 4 screws Chc M4x12. The plate must be
in contact with the exhaust box soldered nuts.
Connect the secondary exhaust port to the customer application.
Connection
A103P
A603P / A1003P
Exhaust connection
DN25 ISO-KF
Several fitting accessories are available in the adixen catalog.
adixen Vacuum Products - Operating instructions - A3P Series