GB 02895 - Edition 02 - Oct 13
3/6
C 90
Use of the serial link
List of controls
Order
Paramater
Response
Action
DLI
XXX
OK
or
ERR2
Program the automatic transmission interval: 1 to 255 sec.
DLR
None
OK
Start up the Data Logger for RS 232 only.
ECH
ON
or
OFF
OK
or
ERR2
Return all characters received on the serial link in the ECHON
case for RS 232 only.
HDR
<val>
(see folio 5/6)
OK
or
ERR3
Change the response heading character.
IDN
None
V X.X
Send back the software version number.
SET
XXYZZZZ
OK
or
ERR2
Parameter setting
Example : # 0 0 0 S E T X X Y Z Z Z Z
XX = 00
: adp power
01
:
reserved
02
:
reserved
03
:
nitrogen
flow
04
:
reserved
05
:
reserved
06
: purge flow prolonged
07
:
full
overhaul
08
: P out maintenance
09
:
reserved
10
:
exhaust
Maintenance
11
:
reserved
12
:
reserved
13
: pressure alert threshold
14
: date setting : Day
15
: date setting : Month
16
: date setting : Year
17
: time setting : Hour
18
: time setting : Minute
19
:
free
20
: setpoint (0) / adp T° alarm (1)
21
:
reserved
22
:
reserved
Y = 0
: alert
1
:
alarm
ZZZZ =
threshold value
SEP
<val>
(see folio 5/6)
OK
or
ERR3
Used to change the separator character between the two
consecutive data items in a response.
SYS
ON
or
OFF
OK
or
ERR2
Start up the pumping unit with SYSON and stop with SYSOFF
(see CTRL order).
CTRL
ON
or
OFF
OK
or
ERR2
Take the pump control with CTRLON and let the pump control
with CTRLOFF.
List of controls
adixen Vacuum Products - Operating instructions - A3P Series