Diagnosis and trouble shooting
INCIDENT
CAUSE
TROUBLESHOOTING
Hand held display module is not
lighting up
Main breaker (RT1) in position
«0»
Set the main break (RT1) to
position «1».
Check the ground fault relay
B 50
.
No main power
Check power supply.
Hand held module display cable
incorrectly connected
Check the connection.
Other problem
Call the customer service center.
Display is showing
XXX LOC.N.2
PUMPING STOPPED
MM/JJ/AA HH:MM
... by pressing on any key ...
XXX LOC.N.2
COMMAND NOT ALLOWED
MM/JJ/AA HH:MM
Pump is used with several hand
held module display
Only one hand held module
display can monitor the pump.
Press on CONTROL to disable
the monitoring on other module
displays.
Press on CONTROL on the
requested module display to
monitor the group.
Wait for the display:
* XXX LOC.N.2
PUMPING STOPPED
MM/JJ/AA HH:MM
Other problem
Call the customer service center.
Message on display
STOCK
Pump in storage mode
Set the main switch to position
«OFF» to switch off the pump.
Switch it on again.
Message on display
REM
In remote mode
no cover plug or bad connection
Check the wiring of the remonte
control connector
B 70
.
Monitoring problem
Call the customer service center.
Display doesn’t work or is showing a message