6/8
D 30
Diagnosis and trouble shooting
GB 04155 - Edition 04 - Oct 13
INCIDENT
CAUSE
TROUBLESHOOTING
Hazard
D17 ROOTS TEMP.
Safety hazard
D40 SAFETY ROO. TEMP.
A103P only
Strap disconnected
Connect the «Motor temp Roots
switch» strap on the connector
D 10
.
A603P and A1003P only
Roots motor temperature
is > 150 °C
Refer to water flow defect
(
see W08
).
Sensor problem
Check the sensor connection on
the «Motor temp. Roots switch»
connector
D 10
.
Other problem
Call the customer service center.
Warning
W29 MAINT. ROOTS
Warning
W30 EXHAUST MAINT.
Warning
W31 ADP LP MAINT.
Warning
W32 ADP MAINT.
The maintenance threshold of a
parameter is reached
Check the maintenance thresholds
C 42
.
Call Manufacturer Support
Service.
Hazard
D34 EM. STOP; ADP. CONT
A603P and A1003P only
EMO cover plug disconnected
Connect the EMO cover plug
B 100
.
Check that there is a strap
between pins 2 and 3 of the
EMO
B 100
.
Other problem
Call the customer service center.