GB 02892 - Edition 03 - Oct 13
1/3
C 43
Starting mode definition
«Hand held display module»
Pump start up
Check that the safety plug is connected
B 80
.
Check that the remote control cover plug is connected
B 70
.
Check that the EMO cover plug is connected
B 100
.
Put the main switch on position
1
.
Connect the hand held display module plug to the hand held display
module connector located on the front or rear panel.
Take the pump control via the
CONTROL
key on the hand held display
module (a star appears on the left side of the screen).
Start the pump by pressing the
START
key on the hand held display
module.
The following operations will perform automatically:
Inlet valves open (if preselected).
Pump starts.
Data capture and sensor are in progress.
Purge valve opens.
Then adjust the gas flow rate using the pressure regulator (as required
by customer application
B 40
).
23 H LOC N2 LS RT
ADP POWER : 1000 W
PRESSURE : 1322 HPA
ADP TEMP : 82°C
Note : Only installed and configured parameters are displayed.
During pump start-up, the temperature is not monitored (no warning, no hazard)
until the pump has reached the programmed temperature value.
Study the preliminary precautions
B 00
.
The performance and the operating safety of the product can only be
guaranteed if it is used according to normal conditions of use.
Do not move a pump when it is in operation.
The pump is tested and filled with a synthetic oil charge at factory. Do not
modify the oil level.
For A603P pump model, don’t forget to check the direction of rotation before
starting up the pump
B 51
.
The pumps are equipped with an exhaust pressure monitoring sensor.
It is the responsability of the user to prevent the risks related to excess
overpressure.
These operating parameters are displayed on the screen.
adixen Vacuum Products - Operating instructions - A3P Series