GB 02893 - Edition 01 - December 05
1/2
C 44
Saving and loading of pump configuration
STORAGE
CONFIG SAVE
CONFIG LOAD
MAINTENANCE
MANAGEMENT
OPERATING TIME
LAST WARNING (S)
CONFIG SAVE
CONFIRM ?
At pump installation or replacement, it may be necessary to copy
the configuration (set of all the parameters and operation setting
programmed by the user) from a pump to another.
Enter the menu by pressing
SET
.
Press the
+
key repeatedly to acces to the
MANAGEMENT
menu.
Validate with the
ENTER
key.
Press the
+
key repeatedly to access the
CONFIG SAVE
menu.
Validate with the
ENTER
key.
The following screen appears.
Validate with the
ENTER
key.
The pump configuration is then transfered into the hand held dislay
module memory.
Return to main menu and exit the setting mode by
SET
.
Disconnect the hand held display module.
Saving the
configuration
The pump configuration is stored in the monitoring system.
Loading and saving the parameters with the pump in operation may result
in a software conflict.
Perform these operations when the pump is switched on but with pumping
stopped.
adixen Vacuum Products - Operating instructions - A3P Series