Setting Measurement Items
142
ZW-8000/7000/5000
User's Manual
Calculates using the measurement value of the last TASK.
The timing to update the measurement value of TASK is shifted one cycle by comparing the update timing of the
measurement result of reference TASK.
Example)
TASK1: Height (Edge1)
TASK2: Height (Edge1)
TASK3: Calculation (X: TASK1, Y: TASK2, M: 1.9, K: 0.0)
Measurement count
TASK1
TASK2
TASK3
1st
0.200
0.100
0.000
2nd
0.310
0.110
0.100 *Calculates using 1st result of TASK1 and TASK2.
3rd
0.420
0.120
0.200 *Calculates using 2nd result of TASK1 and TASK2.
4th
0.530
0.130
0.300 * Calculates using 3rd result of TASK1 and TASK2.
Important
Содержание ZW-7000 Series
Страница 22: ...20 ZW 8000 7000 5000 User s Manual Index 329 Revision History 332 ...
Страница 23: ...ZW 8000 7000 5000 User s Manual 1 21 ...
Страница 54: ...Basic Operations for Displaying Measurement Results 52 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 96: ...Saving a project 94 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 192: ...Saving The Settings 190 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 280: ...Connecting by No protocol Communications 278 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 281: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 280 8 2 Troubleshooting 282 ...
Страница 288: ...Troubleshooting 286 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 289: ...9 APPENDICES APPENDICES 9 1 Specifications and External Dimensions 288 9 2 Laser safety 323 9 3 Firmware Update 324 ...
Страница 330: ...Firmware Update 328 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 333: ...Index ZW 8000 7000 5000 User s Manual 331 9 APPENDICES ...
Страница 335: ......