4 - 8
LLE11ARDA001A/E
4.3.1.2 Error 102-Finishing was not completed is displayed.
Is the symptom improved?
Return the system to NIDEK.
No
Completion
Yes
Dress the wheel.
(See 7.2.3.)
Is the maximum processing rotation number 40?
Select 40 of the maximum processing rotation number.
No
Yes
Check the operation of the Y axis by hand when the X
axis of the carriage is located in the finishing position.
Do the cables interfere with the mechanical parts?
Place the cables that interfere with the mechanical
parts.
No
Yes