4 - 17
LLE11ARDA001A/E
4.3.3.3 403-Y Initialize error is displayed.
Is the voltage between 1-2, 3-4, 5-6, 7-8 pins of the
P909(J9) connector on the BA09 board (40350-BA09)
DC+24V?
Does the CA44 cable (40350-CA44) break?
Does the CA16 cable (40350-CA17) break?
Does the alfa step motor (40347-E030) of the Y axis
AS (40350-5300) break?
Replace the alfa step motor driver (40350-E030) for Y
axis with the new one.
(See 6.23.2.)
Is the voltage between 1-5 pins of the P208(J8)
connector on the BA02 board (40350-BA02) DC+5V?
Replace the BA09 board (40350-BA09) with the new
one.
(See 6.1.9.)
Replace the CA44 cable (40350-CA44) with the new
one.
Replace the alfa step motor (40347-E030) of the Y axis
AS (40350-5300) with the new one.
(See 6.22.2.)
Replace the CA36 cable (40350-CA36) with the new
one.
Yes
Yes
No
No
Yes
No
Is the symptom improved?
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
Yes
No
Replace the CA17 cable (40350-CA16) with the new
one.
No
Yes
No
Yes
Is the voltage between GND-2 pins of the P401(J1)
connector on the BA04 board (40350-BA04) as
described below when shading the sensor of the CA36
cable (40350-CA36)?
The sensor is shaded.
:
DC 0V
The sensor is not shaded.
:
DC+5V
Completion
Yes
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
No
Does the CA20 cable (40350-CA20) break?
Replace the CA20 cable (40350-CA20) with the new
one.
No
Yes
Is the voltage between GND-1 pin of the P402(J4)
connector on the BA04 board (40350-BA04) DC+5V?
Yes
Replace the BA04 board (40350-BA04) with the new
one.
(See 6.1.4.)
No
Содержание ME-1000
Страница 8: ...1 2 LLE11ARDA001A E...
Страница 12: ...3 2 LLE11ARDA001A E...
Страница 120: ...7 34 LLE11ARDA001A E...
Страница 121: ...8 1 LLE11ARDA001A E 8 SUPPLEMENT 8 1 Wiring Diagram...
Страница 122: ...8 2 LLE11ARDA001A E 8 2 Connector Cable...
Страница 123: ...8 3 LLE11ARDA001A E...
Страница 125: ...8 5 LLE11ARDA001A E 8 4 Labels Top view Processing chamber...
Страница 126: ...8 6 LLE11ARDA001A E Rear view or...