4 - 29
LLE11ARDA001A/E
4.3.6.1 701-Cover open error is displayed.
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
Is the voltage of the P206(J6) on the BA02 board
(40350-BA02) as described below?
1-9(2-10) pins
:
DC+24V
7-9(8-10) pins
:
DC+ 5V
Does the CA15 cable (40350-CA15) break?
D o e s t h e p u l s e m o t o r ( 4 0 3 5 0 - E 0 2 0 ) f o r t h e
electrically-operated cover break?
Is there a continuity between the 1-4 pins of the P803
connector when turning ON the switch of the CA33
cable (40350-CA33)?
Is the voltage between the 1-4 pins of the P803
connector for the CA33 cable (40350-CA33) as
described below?
The cover is open.
:
There is a continuity.
The cover is close.
:
There is no continuity.
Does the CA12 cable (40350-CA12) break?
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
Replace the CA15 (40350-CA15) with the new one.
Replace the pulse motor (40350-E020) for the
electrically-operated cover with the new one.
(See 6.27.)
Adjust the position of the switch for the CA33 cable
(40350-CA33) with the new one.
(See 6.34.)
Replace the CA12 cable (40350-CA12) with the new
one.
No
No
Yes
Yes
No
No
Yes
Yes
Yes
No
Yes
Replace the CA33 cable (40350-CA33) with the new
one.
(See 6.34.)
No
Is the symptom improved?
Replace the BA01 board (40350-BA01) with the new
one.
(See 6.1.1.)
Completion
No
Yes
Yes
Содержание ME-1000
Страница 8: ...1 2 LLE11ARDA001A E...
Страница 12: ...3 2 LLE11ARDA001A E...
Страница 120: ...7 34 LLE11ARDA001A E...
Страница 121: ...8 1 LLE11ARDA001A E 8 SUPPLEMENT 8 1 Wiring Diagram...
Страница 122: ...8 2 LLE11ARDA001A E 8 2 Connector Cable...
Страница 123: ...8 3 LLE11ARDA001A E...
Страница 125: ...8 5 LLE11ARDA001A E 8 4 Labels Top view Processing chamber...
Страница 126: ...8 6 LLE11ARDA001A E Rear view or...