![Nidek Medical ME-1000 Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/nidek-medical/me-1000/me-1000_service-manual_1688027065.webp)
6 - 11
LLE11ARDA001A/E
6.10 Replacing the Spindle AS
Replacement part: 40350-1400
1. Remove the front cover (40350-M220) and
side cover L (40350-M225). (See the section
5.)
2. Slide the right and left shafts (40350-M505,
M506) in the back.
3. Remove the wheel.
(See 6.2.)
4. Disconnect the connector of the brushless
motor (40350-E010).
5. Unscrew SB6×20 (n=4) and remove the
spindle AS (40350-1400).
* Pay attention to avoid catching the spindle
AS in the guard inside of the processing
chamber.
6. Reassemble the parts in reverse order.