![Nidek Medical ME-1000 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/nidek-medical/me-1000/me-1000_service-manual_1688027021.webp)
4 - 9
LLE11ARDA001A/E
4.3.2.1 201-Feeler off error is displayed.
Replace the lens with the new one, or change the
layout.
Is the lens size less than the processing range?
Does the left feeler (40350-M435) or the right feeler
(40350-M436) wear out remarkably?
Does the rotary encoder (40350-E035) of the left lens
measuring AS (40350-4100) break?
Replace the left feeler (40350-M435) or the right feeler
(40350-M436) with the new one.
(See 6.18 and 6.19.)
No
Yes
No
Yes
Replace the rotary encoder (40350-E035) of the left
lens measuring AS (40350-4100) with the new one.
(See 6.21.1.)
No
Yes
Does the CA13 cable (40350-CA13) break?
Replace the CA13 cable (40350-CA13) with the new
one.
No
Yes
Does the DC motor (40350-E047) of the left lens
measuring AS break?
Replace the DC motor (40350-E047) of the left lens
measuring AS (40350-4100) with the new one.
(See 6.20.1.)
No
Yes
Does the DC motor (40350-E047) of the right lens
measuring AS (40350-4200) break?
Replace the DC motor (40350-E047) of the right lens
measuring AS (40350-4200) with the new one.
(See 6.20.2.)
No
Yes
Does the right feeler AS (40350-4220) move smoothly
when moving it to the right and left by hand?
Replace the right lens measuring AS (40350-4200)
with the new one.
(See 6.9.2.)
Yes
No
Does the left feeler (40350-4120) move smoothly when
moving it to the right and left by hand?
Replace the left lens measuring AS (40350-4100) with
the new one.
(See 6.9.1.)
Yes
No
Is there a continuity between the connectors below on
the BA07 board (40350-BA07) of the left lens
measuring AS (40350-4100)?
J1 connector
J3 connector
Between 7
-
1 pins
Between 8
-
2 pins
Replace the BA07 board (40350-BA07) of the left lens
measuring AS (40350-4100).
(See 6.1.7.1.)
Yes
No
Is there a continuity between the connectors below on
the BA07 board (40350-BA07) of the right lens
measuring AS (40350-4200)?
J1 connector
J3 connector
Between 7
-
1 pins
Between 8
-
2 pins
Replace the BA07 board (40350-BA07) of the right
lens measuring AS (40350-4200) with the new one.
(See 6.1.7.2.)
Yes
No
Содержание ME-1000
Страница 8: ...1 2 LLE11ARDA001A E...
Страница 12: ...3 2 LLE11ARDA001A E...
Страница 120: ...7 34 LLE11ARDA001A E...
Страница 121: ...8 1 LLE11ARDA001A E 8 SUPPLEMENT 8 1 Wiring Diagram...
Страница 122: ...8 2 LLE11ARDA001A E 8 2 Connector Cable...
Страница 123: ...8 3 LLE11ARDA001A E...
Страница 125: ...8 5 LLE11ARDA001A E 8 4 Labels Top view Processing chamber...
Страница 126: ...8 6 LLE11ARDA001A E Rear view or...