4 - 27
LLE11ARDA001A/E
4.3.5.2 Drill/Cutter rotation error is displayed (only for the former
versions than V1.10).
Replace the CA23 cable (40350-CA23) with the new
one.
Does the CA23 cable (40350-CA23) break?
Does the DC motor with encoder (40350-E026) for
grooving & drilling AS break?
Does the CA12 cable (40350-CA12) break?
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
Is the symptom improved?
Completion
Replace the DC motor with encoder (40350-E026) for
grooving & drilling AS with the new one.
(See 6.26.2.)
Replace the BA01 board (40350-BA01) with the new
one.
(See 6.1.1.)
Yes
No
Yes
No
Yes
No
Replace the CA12 cable (40350-CA12) with the new
one.
Replace the CA23 cable (40350-CA24) with the new
one.
Does the CA23 cable (40350-CA24) break?
Yes
No
No
Yes