4 - 24
LLE11ARDA001A/E
4.3.3.13 413-SFB Initialize error is displayed.
Replace the CA21 cable (40350-CA21) with the new
one.
Does the CA21 cable (40350-CA21) break?
Is there a continuity between the following connectors
on the BA05 board (40350-BA05)?
P501(J1) connector - P505(J5) connector
1
-
1 pins
2
-
2 pins
3
-
3 pins
4
-
4 pins
5
-
5 pins
6
-
6 pins
P501(J1) connector - P506(J6) connector
7(8)
-
1 pins
11
-
2 pins
Is the voltage between the 2nd pin-GND of the
P506(J6) connector on the BA05 board (40350-BA06)
as described below when shading the sensor of the
CA39 cable (40350-CA39)?
The sensor is shaded.
:
DC 0V
The sensor is not shaded.
:
DC+5V
Does the CA12 cable (40350-CA12) break?
Replace the BA05 board (40350-BA05) with the new
one.
(See 6.1.5.)
Replace the CA39 cable (40350-CA39) with the new
one.
Replace the CA12 cabel (40350-CA12) with the new
one.
No
Yes
Yes
Yes
No
No
No
Yes
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
Replace the BA01 board (40350-BA01) with the new
one.
(See 6.1.1.)
Is the symptom improved?
No
Yes
Is the voltage between the 1-9(2-10) pins on the BA02
board (40350-BA02) DC+24V, and the one between
the 7-9(8-10) pins DC+5V?
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
Yes
No
Completion
Содержание ME-1000
Страница 8: ...1 2 LLE11ARDA001A E...
Страница 12: ...3 2 LLE11ARDA001A E...
Страница 120: ...7 34 LLE11ARDA001A E...
Страница 121: ...8 1 LLE11ARDA001A E 8 SUPPLEMENT 8 1 Wiring Diagram...
Страница 122: ...8 2 LLE11ARDA001A E 8 2 Connector Cable...
Страница 123: ...8 3 LLE11ARDA001A E...
Страница 125: ...8 5 LLE11ARDA001A E 8 4 Labels Top view Processing chamber...
Страница 126: ...8 6 LLE11ARDA001A E Rear view or...