4 - 13
LLE11ARDA001A/E
Does the CA12 cable (40350-CA12) break?
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
Replace the CA12 cable (40350-CA12) with the new
one.
No
Yes
Yes
Is the symptom improved?
Replace the BA01 board (40350-BA01) with the new
one.
(See 6.1.1.)
Completion
No
Yes
Is the symptom improved?
Completion
No
Yes
Is the voltage between 6-4 pins of the P705(J5)
connector on the BA07 board (40350-BA07) of the
right lens measuring AS (40350-4200) DC+5V?
Adjust the position of the shading plate (40350-M424).
(See 7.16.)
No
Is the voltage between 6-4 pins of the P705 (J5)
connector of the left lens measuring AS (40350-4100)
DC+5V?
Adjust the position of the shading plate (40350-M428).
(See 7.15.)
Yes
No