8 - 7
LLE11ARDA001A/E
8.5 Error message
No.
Error indication
Contents
Causes and measures
101
101-Roughing was not
completed
Roughing is not completed.
Insufficient dressing of the Glass
wheel
102
102-Finishing was not
completed
Finishing is not completed.
Insufficient dressing of the wheel
(This error message is displayed
when the rotating numbers of the
θ
axis achieves unlikely one.)
201
201-Feeler off error
Feeler is off when starting the
lens measurement.
The lens diameter is insufficient.
202
202-There is no lens
Feeler is off.
Lens is not attached.
203
203-Feeler lift error
Feeler lifts up.
Grind the cataract lens, bifocal
lens and such in EX mode.
204
204-Front & rear data crossed
Measured data is improper.
(Front and rear data was
crossed.)
Failure of the encoder (The thin
lens is deformed.)
205
205-Start & end measure data
error
Measured data at the start point
does not match to the one at the
end point.
Failure of the encoder (Check the
deformation of the thin lens and
attachment of the feelers)
401
401-XYR Initialize error
The X, Y and R axes are not
initialized.
Check the driving transferring unit
of each origin sensor and gears.
402
402-X Initialize error
The X axis is not initialized.
Same as above
403
403-Y Initialize error
The Y axis is not initialized.
Same as above
404
404-R Initialize error
The R axis is not initialized.
Same as above
405
405-XY Initialize error
The X and Y axes are not
initialized.
Same as above
406
406-XR Initialize error
The X and R axes are not
initilized.
Same as above
407
407-YR Initialize error
The Y and R axes are not
initialized.
Same as above
408
408-Front feeler Initialize error
Front of the F axis is not
initialized.
Same as above
409
409-Rear feeler Initialize error
Rear of the F axis is not
initialized.
Same as above
410
410-Drill ZR Initialize error
Drill Z and R axes are not
initialized.
Same as above
411
411-Drill Z Initialize error
Drill Z axis is not initialized.
Same as above
412
412-Drill R Initialize error
Drill R axis is not initialized.
Same as above
413
413-SFB Initialize error
The SFB arm is not initialized.
Same as above
501
501-Chuck origin error
Chuck origin sensor does not
respond.
Same as above
601
601-Main wheel was not rotate The main wheel does not rotate.
Check the LED indication of the
inverter.
603
603-SFB wheel was not rotate The SFB wheel does not rotate.
Same as above
701
701-Cover open error
Cover error (OPEN)
The cover does not open by
pressing the Cover button after the
processing.
702
702-Cover close error
Cover error (CLOSE)
The cover does not close by
pressing the Start and Cover
buttons.
Содержание ME-1000
Страница 8: ...1 2 LLE11ARDA001A E...
Страница 12: ...3 2 LLE11ARDA001A E...
Страница 120: ...7 34 LLE11ARDA001A E...
Страница 121: ...8 1 LLE11ARDA001A E 8 SUPPLEMENT 8 1 Wiring Diagram...
Страница 122: ...8 2 LLE11ARDA001A E 8 2 Connector Cable...
Страница 123: ...8 3 LLE11ARDA001A E...
Страница 125: ...8 5 LLE11ARDA001A E 8 4 Labels Top view Processing chamber...
Страница 126: ...8 6 LLE11ARDA001A E Rear view or...