4 - 26
LLE11ARDA001A/E
Does the CA10 cable (40350-CA10) break?
Does the E059 cable (40347-E059) break?
Does the E060 cable (40347-E060) break?
Replace the spindle motor AS (40350-1410) with the
new one.
(See 6.24.)
Is the symptom improved?
Replace the inverter for the BL motor (40350-
E031(100V) and 40350-E032(200V) with the new one.
(See 6.25.)
Is the symptom improved?
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
Replace the CA10 cable (40350-CA10) with the new
one.
Replace the E059 cable (40347-E059) with the new
one.
Completion
Completion
No
No
Yes
Yes
Replace the E060 cable (40347-E060) with the new
one.
No
Yes
No
Yes
No
Yes
Turn OFF the power. Then, turn ON the power again
after 20 or more seconds.
Is the symptom improved?
Completion
No
Yes
4.3.5.1 Error 601-Main wheel was not rotate is displayed.