7 - 9
LLE11ARDA001A/E
7. Perform the adjustment by following the
procedures below when the measuring track
is not parallel to the lens edge.
* Recognize the inner line of the track as the
measuring track.
1) Establish the Parameter Setting mode.
(See 7.1.2.)
2) Press Feeler AXIS.
3) Change the parameter settings so that the
measuring track is parallel to the lens edge.
a. The track slants to the right.: decrease the
value.
b. The track slants to the left.: increase the value.
* Enter 1.0 when the measuring track slants to
the right in 1º toward the lens edge.
8. After the parameter settings have been
changed, press BACK to display the
Parameter Setting screen.
9. Press BACK to display the layout screen.
4QWIJ5K\G
(KPKUJ5K\G
2QNKUJ5K\G
2QNKUJ&KHH
#:+5
5CHGV[$GXGN
)TQQXG&TKNN
Adjustment
010'
(GGNGT*GKIJV
(GGNGT#:+5
$GXGN2QUKVKQP
$GXGN2QUKVKQP)NCUU
$GXGN2QUKVKQP2QNKUJ
2QNKUJ9JGGN*GKIJV
&TKNN#FLWUVOGPV/QFG
010'
(GGNGT*GKIJV
(GGNGT#:+5
$GXGN2QUKVKQP
$GXGN2QUKVKQP)NCUU
$GXGN2QUKVKQP2QNKUJ
2QNKUJ9JGGN*GKIJV
&TKNN#FLWUVOGPV/QFG
BACK
Содержание ME-1000
Страница 8: ...1 2 LLE11ARDA001A E...
Страница 12: ...3 2 LLE11ARDA001A E...
Страница 120: ...7 34 LLE11ARDA001A E...
Страница 121: ...8 1 LLE11ARDA001A E 8 SUPPLEMENT 8 1 Wiring Diagram...
Страница 122: ...8 2 LLE11ARDA001A E 8 2 Connector Cable...
Страница 123: ...8 3 LLE11ARDA001A E...
Страница 125: ...8 5 LLE11ARDA001A E 8 4 Labels Top view Processing chamber...
Страница 126: ...8 6 LLE11ARDA001A E Rear view or...