5-16
5.4 Display Electronic Flowmeter
Gas flow can be displayed either in a standard-resolution mode or high-resolution mode.
These two resolution modes vary in scale and accuracy.
Switchover between the standard-resolution mode and the high-resolution mode can be
performed manually based on gas flow. The default is high-resolution mode.
The scale range of the standard-resolution mode is 0 to 10 L/min and that of the
high-resolution mode 0 to 6 L/min.
To select the electronic flowmeter with desired resolution, select the electronic flowmeter
area to access the [
Display Selection
] menu.
When spirometry loop and gas modules are configured, you can select to display either
spirometry loop or electronic flowmeter. In this case, select the electronic flowmeters area to
open the [
Display
Selection
] menu and then select the desired spirometry loop or electronic
flowmeter.
5.5 Spirometry Loop
Spirometry loops reflect patient lung function and ventilation as well, such as compliance,
over-inflation, breathing system leak and airway blockage.
The system provides two spirometry loops: P-V (Paw-volume) loop and F-V (flow-volume)
loop. Only one loop is displayed at a time: P-V loop or F-V loop. To switch over between the
two loops, select spirometry loop area and select the desired loop. The scales of volume flow
and Paw are adjusted automatically. The following figures show an F-V loop and a P-V loop.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...