4-20
In PSV mode, you also need to set:
[
Trigger Level
]
1. Select
the
[
Vent Mode
] shortcut key
→
[
PSV >>
]
→
[
Trigger Level
].
2. Select
[
Pressure
] or [
Flow
] for trigger type.
3. Turn the control knob to set [
Trigger Level
] to the appropriate value.
4. Push the control knob to confirm the setting.
5. Select
[
Ok
]
to activate the current setting.
6. To cancel the current setting and exit the current menu, select [
Cancel
],
or push the
Normal Screen key.
[
PSV
Insp Termination Level
]
Inspiration termination level refers to the percentage of inspiration flow to the maximum
inspiration flow during inspiration in the PSV mode.
To set [
PSV
Insp Termination Level
], do as follows:
1. Select
the
[
Vent Mode
] shortcut key
→
[
PSV >>
]
→
[
PSV
Insp Termination
Level
].
2. Push the control knob and turn it to set [
PSV Insp Termination Level
]
to the
appropriate value.
3. Push the control knob to confirm the setting.
4. Select
[
Ok
]
to activate the current setting.
5. To cancel the current setting and exit the current menu, select [
Cancel
],
or push the
Normal Screen key.
[
Backup Mode Active
]
When PSV mode is applied alone, the PCV backup mode is available. If within the preset
time (Backup Mode Active), no spontaneous breathing occurs or the spontaneous breathing is
not strong enough to reach Trigger Level, the PCV backup mode is activated automatically
when the time for Backup Mode Active is up to enable mechanical ventilation forcibly.
To set [
Backup Mode Active
], do as follows:
1. Select
the
[
Vent Mode
] shortcut key
→
[
PSV >>
]
→
[
Backup Mode Active
].
2. Push the control knob and turn it to set [
Backup Mode Active
]
to the appropriate value.
3. Push the control knob to confirm the setting.
4. Select
[
Ok
]
to activate the current setting.
5. To cancel the current setting and exit the current menu, select [
Cancel
],
or push the
Normal Screen key.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...