D-9
Flowmeter
Hardware Error 06
H
Flash selftest error
Flowmeter
Hardware Error 07
H
Watchdog selftest error
Flowmeter Cal.
Data Error 01
H
Air, O
2
and N
2
O data empty
Flowmeter Cal.
Data Error 02
H
Air, O
2
and N
2
O data error
Flowmeter Comm
Error
H
XX module failed to
communicate with the main
system.
Flowmeter Comm
Stop
H
The electronic flowmeter
failed to communicate with
the main system normally.
N2O Flow Too
High
L
The N
2
O flow control is
turned to set the flow too
high.
O2 Flow Too High
L
The O
2
flow control is turned
to set the flow too high.
Air Flow Too High
L
The Air flow control is turned
to set the flow too high.
Turn the flow control to
keep the flow within 10
L/min.
O2-N2O Ratio
Error
H
Incorrect O2-N2O ratio. Contact your service personnel.
Flowmeter Zero
Failed
L
Board or 3-way valve malfunction. Contact your service
personnel.
AG Init Error
H
The AG module was installed improperly or
malfunctioned.
AG Cal. Failed
H
The AG module calibration failed.
AG Comm Stop
H
AG module malfunction or communication failure
AG Airway
Occluded
H
The actual pump rate of the AG module is less than 20
mL/min for more than one second.
AG Comm Error
H
AG module communication failure
AG Hardware Error M AG module hardware error
AG Selftest Error
H
Module fault or communication failure between the
module and anesthesia machine. Re-plug the module,
restart the anesthesia machine, or try to plug the module
into another anesthesia machine.
AG
module
AG Hardware
Malfunction
H
AG module hardware malfunction. The AG module enters
Standby and measurement stops. Remove the AG module
and contact your service personnel.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...