D-5
Device Fault,
Ventilate Manually
H
Equipment malfunction. Mechanical ventilation and
monitoring did not work. Use manual ventilation mode to
help the patient breathe and restart the anesthesia machine.
If the alarm cancels, restart mechanical ventilation.
Ventilator
Hardware Error 01
H
CPU error.
Ventilator
Hardware Error 02
H
RAM error.
Ventilator
Hardware Error 03
H
ROM error.
Ventilator
Hardware Error 04
H
Watchdog error.
Ventilator
Hardware Error 05
H
EEPROM error.
Ventilator
Hardware Error 06
H
Internal AD error.
Ventilator
Hardware Error 07
H
External AD error.
Ventilator
Hardware Error 08
H
5 V power error.
Ventilator
Hardware Error 09
H
12 V power error.
Ventilator
Hardware Error 11
H
Safety valve control
failure by the auxiliary
control board.
Ventilator
Hardware Error 12
H
Safety valve control
failure by the main
control board.
Unreliable monitoring. Use
manual ventilation mode to help
the patient breathe Contact your
service personnel.
Auxi Ctrl Module
Error
H
No message of test completion is received after 10-second
wait during the auxiliary control board pressure
effectiveness and safety valve control effectiveness test.
No message of zeroing completion is received after
10-second wait when the auxiliary control board is
notified to zero. Use manual ventilation mode to help the
patient breathe Contact your service personnel.
Ventilator Comm
Error
H
The ventilator module failed to communicate with the
main system. Contact your service personnel.
Ventilator
Ventilator Comm
Stop
H
The ventilator module failed to communicate with the
main system normally. Unreliable monitoring. Use manual
ventilation mode to help the patient breathe Contact your
service personnel.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...