2-11
WARNING
z
Connect to the AC mains in compliance with B.3 Power Requirements. Failure to
do so may cause damage to the equipment or affect its normal operation.
z
Make sure that the jacket on the electrical outlet is already fixed to avoid power
cord off during surgery.
NOTE
z
If the equipment cannot be powered by the AC mains, check if the fuse inside the
electrical outlet is normal. If AC mains supply still fails after the fuse is replaced,
contact the service personnel.
z
When the auxiliary electrical outlet does not work normally, check if the
corresponding fuse is burned.
z
Equipment connected to the auxiliary electrical outlet shall be authorized.
Otherwise, leakage current above the allowable limit will result, which may
endanger the patient or operator, and damage the anesthesia machine or externally
connected equipment. When the anesthesia machine is configured with only one
auxiliary electrical outlet, this electrical outlet is only used for connecting the
adapter for Desflurane vaporizer. When the anesthesia machine is configured with
multiple auxiliary electrical outlets, the equipment connected shall comply with the
voltage and current specifications of the auxiliary electrical outlets.
z
All analog or digital products connected to this system must be certified passing the
specified IEC standards (such as IEC 60950 for data processing equipment and
IEC 60601-1 for medical electrical equipment). All configurations shall comply
with the valid version of IEC 60601-1-1. The personnel who are responsible for
connecting the optional equipment to the I/O signal port shall be responsible for
medical system configuration and system compliance with IEC 60601-1-1 as well.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...