CO2 Monitoring
8.1 Introduction
CO
2
monitoring is a continuous, non-invasive technique for determining the concentration of
CO
2
in the patient’ airway by measuring the absorption of infrared (IR) light of specific
wavelengths. The CO
2
has its own absorption characteristic and the amount of light passing
the gas probe depends on the concentration of the measured CO
2.
When a specific band of IR
light is passed through respiratory gas samples, some of IR light will be absorbed by the CO
2
molecules. The amount of IR light transmitted after it has been passed through the respiratory
gas sample is measured with a photodetector. From the amount of IR light measured, the
concentration of CO
2
is calculated.
There are two methods for measuring CO
2
in the patient’s airway:
1. Mainstream
measurement
Uses a CO
2
sensor attached to an airway adapter directly inserted into the patient’s
breathing system.
2. Sidestream/microstream
measurement
Samples expired patient gas at a constant sample flow from the patient’s airway and
analyzes it with a CO
2
sensor built into the CO
2
module.
The measurement provides:
1. CO
2
waveform.
2. End-tidal
CO
2
(EtCO
2
) value: the CO
2
value measured at the end of the expiration
phase.
3. Fraction of inspired CO
2
(FiCO
2
): the CO
2
value measured during inspiration.
NOTE
z
Perform CO
2
monitoring when using this anesthesia machine to ensure patient
safety. If your anesthesia machine is not configured with CO
2
module, use the
anesthesia machine with
CO
2
monitoring function in compliance with the relevant
international standard to perform CO
2
monitoring.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...