7-2
7.2 Maintenance Schedule
NOTE
z
These schedules are the minimum frequency based on typical usage of 2000 hours
per year. You should service the equipment more frequently if you use it more than
the typical yearly usage.
Minimum
frequency
Maintenance
Daily
Clean the external surfaces.
21%O2 calibration (O
2
sensor in breathing system).
Biweekly Drain
the
vaporizers.
Monthly
100% O
2
calibration
(breathing system O
2
sensor).
Clear water built up inside the waterstraps of CO2 module and AG module.
During cleaning
and setup
Inspect the parts and seals for damage. Replace or repair as necessary.
Annually
Replace the seal on the vaporizer manifold and that on the breathing system
port. Contact us for details.
CO2 module calibration.
AG module calibration.
Every three
years
Replace the built-in lithium-ion batteries. Contact us for details.
As necessary
Before installing the cylinder, use a new cylinder gasket on cylinder yoke.
Empty the water collection cup If there is water built up in it.
Replace the sodalime in the canister if sodalime color change is detected.
Replace the O2 sensor if a great deviation of the measured value by the O2
sensor occurs and the problem persists after multiple calibrations.
Replace the flow sensor if the seal for the flow sensor is damaged, the
membrane inside the flow sensor is cracked or distorted, or the flow sensor is
cracked or distorted.
Replace the transfer tube if it is damaged.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...