6-13
6.9.2 Test the O2 Concentration Monitoring and Alarms
NOTE
z
This test is not required if no O
2
sensor is configured.
1. Set the bag/mechanical ventilation switch to the bag
position.
2. Remove the O
2
sensor. After two to three minutes, make sure that the sensor measures
approximately 21% O
2
in room air.
3. Select
the
[
Alarm Setup
] shortcut key and then [
Ventilator >>
]. Set the FiO
2
low alarm
limit to 50%.
4. Make sure that a low FiO
2
alarm occurs.
5. Set the FiO
2
low alarm limit back to a value less than the measured FiO
2
value and make
sure that the alarm cancels.
6. Put the O
2
sensor back in the breathing system.
7. Select
the
[
Alarm Setup
] shortcut key and then [
Ventilator >>
]. Set the FiO
2
high
alarm limit to 50%.
8. Connect the manual bag to the manual bag port. Push the O
2
flush button
to fill the
manual bag. Make sure that the sensor measures approximately 100% O
2.
9. Make sure that a high FiO
2
alarm occurs.
10. Set the FiO
2
high alarm limit to 100% and make sure that the alarm cancels.
6.9.3 Test the Low Minute Volume Alarm
1. Make sure that MV alarm is switched on.
2. Select
the
[
Alarm Setup
] shortcut key and then [
Ventilator
>>]. Set the MV low alarm
limit to 8.0 L/min.
3. Make sure that a low MV alarm occurs.
4. Select
the
[
Alarm
Setup
] shortcut key and then [
Ventilator
>>]. Set the MV low alarm
limit to 2.0 L/min.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...