5-10
4. Select to exit the current menu.
5. Set waveform scale. The Paw waveform scale is automatically adjusted based on the set
Plimit. You can set the Paw waveform scale appropriately by setting Plimit.
5.3.4.4 Set Paw Unit
1. Select
the
[
Maintenance
] shortcut key and select [
User Maintenance >>
].
2. Select
[
Paw Unit
] and toggle between [
cmH2O
], [
hPa
]
and [
mbar
].
3 Select to exit the current menu.
5.3.4.5 Review Ppeak Trend
For details about reviewing Ppeak trend, refer to
12 Trend and Logbook.
5.3.5 Tidal Volume Monitoring
NOTE
z
The tidal volume marked on the bellows housing is only a rough indicator. It may
be inconsistent with the tidal volume actually measured. This is a normal
phenomenon.
z
As required by the relevant international rules and regulations, tidal volume
monitoring needs to be performed when the anesthesia machine is used on the
patient. If your anesthesia machine is not configured with such monitoring
function, use a qualified monitor for tidal volume monitoring.
5.3.5.1 Display Tidal Volume and Breath Rate Parameters
If your anesthesia machine is configured with CO2 or AG module, tidal volume and breath
rate related parameters are displayed as shown below.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...