B
-5
O2 flush
35 to 75 L/min
Flowmeter
Electronic flowmeters
Air range: 0 to 10 L/min
O2 range: 0 to 10 L/min
N2O range: 0 to 10 L/min
Accuracy: < ±10% of the indicated value (under 20
℃
and 101.3 kPa, for flow between 10% of full scale or 300
mL/min (whichever is greater) and full scale)
Total flowmeter
Type: Rotameter
Range: 0 to 10 L/min
Accuracy: < ±10% of the indicated value (under
20
℃
and 101.3 kPa, for flow between 10% of full scale or 300
mL/min (whichever is greater) and full scale (calibrated at 100%
O2))
Auxiliary O2 supply
Gas supply: O2 in the system
Flow: 0 to 10 L/min
Accuracy: ±5% of full range (under 20
℃
and
101.3 kPa, for flow between 10% of full scale or 300 mL/min
(whichever is greater) and full scale (calibrated at 100% O2));
pressure compensation not provided
O2-N2O link system
Type
Mechanical proportion control device
Range
O2 concentration not lower than 25%
B
.6 Breathing System Specifications
System leakage and system compliance
System leakage
Not greater than 150 mL/min at 3 kPa
System compliance
≤
4 mL/100Pa in adult mode
Sodalime canister
leakage
Not greater than 50 mL/min at 3 kPa
APL valve leakage
Not greater than 50 mL/min
(
The scale of APL valve is 75 cmH
2
O
)
CO
2
absorber canister
Volume
Approximately 1350 ml
Water collection cup
Type
Can be disassembled independently
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...