4-6
To ensure the set tidal volume gas delivery, the ventilator adjusts gas flow based on the
measured inspiratory volume, dynamically compensates for the loss of tidal volume arising
from breathing system compliance and system leakage and eliminates the effect of fresh gas
as well. This is called tidal volume compensation.
In the VCV
mode, if tidal volume compensation is turned off or failed, the anesthesia
machine can continue delivering gas stably but cannot compensate for the effects of fresh gas
flow and breathing system compliance losses.
4.4.3.2 Waveforms
The following figures show the Paw waveform and flow waveform in the VCV
mode.
Generally, in the VCV mode, the flow waveform is at a constant flow during inspiration and
the Paw waveform rises in the same period.
4.4.3.3 Start VCV Mode
1. Select
the
[
Vent Mode
] shortcut key to open the [
Vent Mode Setup
] menu.
2. Select
[
VCV
] in the [
Vent Mode Setup
] menu.
3. After confirming the selection, the[
TV
] shortcut key (the first key from the left in the
parameter setup shortcut keys area) is highlighted.
4. Make sure that TV is appropriately set for the patient. Push the control knob to confirm
the setting so as to start VCV mode.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...