D-2
TVe Below
Control Range
M
In the VCV mode, TVe is lower than the minimum tidal
volume setting for five continuous breathing cycles. Check
the patient’s condition, pneumatic circuit connection and
flow sensor.
MV Too High
M
MV is higher than the high alarm limit setting. If ventilation
mode is switched or settings for ventilator parameters are
changed, this alarm is disabled temporarily within nine
breathing cycles or one minute (whichever is less) after
being set. Decrease settings for tidal volume or breath rate,
or increase high alarm limit.
MV Too Low
M
MV is lower than the low alarm limit setting. If ventilation
mode is switched or settings for ventilator parameters are
changed, this alarm is disabled temporarily within nine
breathing cycles or one minute (whichever is less) after
being set. Increase settings for tidal volume or breath rate, or
decrease low alarm limit.
Apnea Alarm
M
Two triggering conditions are met simultaneously:
1. Paw is lower than (PEEP+3) cmH
2
O for more than 20
seconds.
2. TVe is lower than 10 ml for more than 20 seconds.
Increase tidal volume and breath rate settings, or apply
manual ventilation.
Volume
Apnea>2 min
H
No breath has been detected within the last 120 seconds.
Check the patient’s condition. Use manual ventilation mode
to help the patient breathe. Check if the tubes fall off.
Rate Too High
L
Rate is higher than the high alarm limit setting. If ventilation
mode is switched or settings for ventilator parameters are
changed, this alarm is disabled temporarily within nine
breathing cycles or one minute (whichever is less) after
being set. Decrease breath rate setting or increase high alarm
limit.
Rate Too Low
L
Rate is lower than the low alarm limit setting. If ventilation
mode is switched or settings for ventilator parameters are
changed, this alarm is disabled temporarily within nine
breathing cycles or one minute (whichever is less) after
being set. Increase breath rate setting or decrease low alarm
limit.
Pressure
Limiting
L
Paw is greater than Plimit. Increase Plimit or decrease TV or
Rate.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...