14-16
All parts of the breathing system can be cleaned and disinfected. The cleaning and
disinfection methods are different for different parts.
You need to select the appropriate method to clean and disinfect the parts based on the actual
situations to avoid cross-contamination.
This table is our recommended cleaning and disinfection methods for all parts of the
breathing system.
Intermediate level disinfection
High level
disinfection
Parts
A* B* C*
Breathing tubes and Y piece
★
★
Breathing mask
★
★
Flow sensor
★
Bellows assembly
★
★
Inspiratory and expiratory check
valves assemblies
★
★
O
2
sensor
★
Canister assembly
★
★
Canister connection block
assembly
★
★
Water collection cup
★
★
Bag arm
★
★
BYPASS assembly
★
★
Breathing system
★
★
Manual bag
★
★
AGSS assembly
★
★
indicates that this disinfection method is applicable.
A*. Clean with a damp cloth soaked in mild detergent and then wipe off the remaining
detergent with a dry lint free cloth.
B*. Flush with water first; then soaked in water and cleaning solution (water temperature
40ºC recommended) for approximately three minutes and wipe with 70% ethanol.
C*. Steam autoclave at maximum 134ºC.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...