14-21
WARNING
z
The end of inspiration/expiration connectors which connects the breathing tube
shall be kept downward to prevent condensed water from entering the breathing
system.
14.3.10 O2 Sensor
WARNING
z
Do not put both of the breathing system and the O
2
sensor in liquid or autoclave
them.
z
Water vapor may condense on the surface of the O
2
sensor, which can result in
invalid O
2
concentration measurement. In this case, you need to take out the O
2
sensor, remove the water condensed on its surface, and reinstall it into the
breathing system.
1. Refer to the methods recommended in the table of
14.3 Clean&Disinfect and Re-install
the Breathing System
to clean and disinfect the O2 sensor.
2. When
the
O
2
sensor is fully dry, refer to
14.2.1 O2 Sensor
to install it in the reverse
order.
14.3.11 AGSS Transfer and Receiving System
14.3.11.1 Filter
Shake the removed filter to get rid of the dust and other contamination until satisfactory
cleaning effect is achieved.
14.3.11.2 Float
Refer to the methods recommended in the table of
14.3 Clean&Disinfect and Re-install the
Breathing System
to clean and disinfect the float.
14.3.11.3 Spoiler
Refer to the methods recommended in the table of
14.3 Clean&Disinfect and Re-install the
Breathing System
to clean and disinfect the spoiler.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...