7-1
7
User Maintenance
7.1 Repair Policy
WARNING
z
Only use lubricants approved for anesthesia or O
2
equipment.
z
Do not use lubricants that contain oil or grease. They burn or explode in high O
2
concentrations.
z
Obey infection control and safety procedures. Used equipment may contain blood
and body fluids.
z
Movable parts and removable components may present a pinch or a crush hazard.
Use care when moving or replacing system parts and components.
Do not use malfunctioning anesthesia machine. Have all repairs and service done by an
authorized service representative. Replacement and maintenance of tube parts listed in this
manual may be undertaken by a competent, trained individual having experience in the repair
of devices of this nature.
After repair, test the anesthesia machine to ensure that it is functioning properly, in
accordance with the specifications.
NOTE
z
No repair should ever be attempted by anyone not having experience in the repair
of devices of this nature.
z
Replace damaged parts with components manufactured or sold by us. Then test the
unit to make sure that it complies with the manufacturer’s published specifications.
z
Contact us for service assistance.
z
For further information about the product, contact us. We can provide documents
about some parts depending on the actual condition.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...