14-18
14.3.4 Breathing Mask
Refer to the methods recommended in the table of
14.3 Clean&Disinfect and Re-install the
Breathing System
to clean and disinfect the breathing mask.
14.3.5 Inspiratory and Expiratory Check Valves Assembly
1. Refer to the methods recommended in the table of
14.3 Clean&Disinfect and Re-install
the Breathing System
to clean and disinfect the inspiratory and expiratory check valves
assembly.
2. Immerse the check valves and their covers in the disinfectant or autoclave them.
Maximum recommended temperature is 134ºC.
3. After they are fully dry, install the inspiratory and expiratory check valves with
reference to
14.2.8 Expiratory Check Valve Assembly
and
14.2.9 Inspiratory Check
Valve Assembly
in the reverse order. Push the check valve into the breathing system and
then turn the valve cover clockwise to tighten it.
WARNING
z
Do not separate the check valve diaphragm from the valve cover.
z
When installing the check valve, depress the valve forcibly to make sure that it is
installed in position.
14.3.6 Bellows Assembly
CAUTION
z
Do not soak the folding bag assembly in warm water and cleaning solution for
more than 15 minutes. Swelling or faster aging can occur.
z
When exposing the folding bag to air dry, hang and outspread it fully to prevent
tackiness.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...