14-19
NOTE
z
Disassemble the bellows assembly before cleaning it. If not, it will take a very
long time to dry.
z
If autoclaving is necessary, assemble the bellows assembly first. Turn over the
bellows assembly to autoclave it.
1. Refer to the methods recommended in the table of
14.3 Clean&Disinfect and Re-install
the Breathing System
to clean and disinfect the bellows assembly.
2. Place the bellows assembly in warm (40
℃
recommended temperature) mild detergent
(such as soap water). Carefully wash the assembly to prevent damage of the parts.
3. Rinse the assembly with clean warm water.
4. Autoclave the cleaned bellows housing. Maximum recommended temperature is 134ºC.
5. Hang the disinfected bellows assembly upside down and dry at a room temperature less
than 70
℃
.
6. Look for damaged parts after the bellows assembly is fully dry. Then install the
assembly with reference to
13.1.5Install the Bellows
.
7. Connect the bellows assembly, ventilator and breathing system.
8. Perform preoperative test before system use. For details, refer to
6.8.1 Bellows Test.
14.3.7 Sodalime Canister
NOTE
z
It is recommended to apply the high level disinfection procedure after the
intermediate level disinfection is completed.
1. Refer to the methods recommended in the table of
14.3 Clean&Disinfect and Re-install
the Breathing System
to clean and disinfect the sodalime canister.
2. Pour the sodalime into the sodalime canister when the canister is fully dry.
3. Refer
to
13.1.8 Install the Sodalime Canister
to install the canister onto the breathing
system.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...