C-3
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip
in U
T
) for 0.5 cycle
40 % U
T
(60 % dip in
U
T
) for 5 cycles
70 % U
T
(30 % dip in
U
T
) for 25 cycles
<5 % U
T
(>95 % dip
in U
T
) for 5 s
<5 % U
T
(>95 % dip
in U
T
) for 0.5 cycle
40 % U
T
(60 % dip in
U
T
) for 5 cycles
70 % U
T
(30 % dip in
U
T
) for 25 cycles
<5 % U
T
(>95 % dip
in U
T
) for 5 s
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the equipment requires
continued operation during
power mains interruptions, it
is recommended that the
equipment be powered from
an uninterruptible power
supply (UPS).
Power frequency
(50/60 HZ)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Guidance and Declaration - Electromagnetic Immunity
The anesthesia machine is suitable for use in the specified electromagnetic environment. The
customer or the user of the anesthesia machine should assure that it is used in such an environment
as described below.
Immunity test IEC60601
test level
Compliance
level
Electromagnetic environment - guidance
Conduced RF
IEC61000-4-6
3 Vrms
150 kHz to 80
MHz
Outside ISM
bands
a
10 Vrms
150 kHz to 80
MHz
In ISM bands
a
3 Vrms
(
V1
)
10Vrms
(
V2
)
Portable and mobile RF communications
equipment should be used no closer to any part of
the system, including cables, than the
recommended separation distance calculated from
the equation appropriate for the frequency of the
transmitter. Recommended separation distances:
P
V
d
⎥⎦
⎤
⎢⎣
⎡
=
1
5
.
3
P
V
d
⎥⎦
⎤
⎢⎣
⎡
=
2
12
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...