4-18
When PSV mode is applied alone, the PCV backup mode is available. If within the preset
time (Backup Mode Active), no spontaneous breathing occurs or spontaneous breathing is not
strong enough to reach Trigger Level, the PCV backup mode is activated automatically when
the time for Backup Mode Active is up to enable mechanical ventilation forcibly.
The PSV mode can be used jointly with SIMV-VC or SIMV-PC.
4.4.6.2 Waveforms
The following figures show the Paw waveform and flow waveform in the PSV
mode.
4.4.6.3 Start PSV Mode
1. Select
the
[
Vent Mode
] shortcut key to open the [
Vent Mode Setup
] menu.
2. Select
the
[
Vent Mode
] shortcut key and then [
PSV >>
] to open the
[
PSV Setup
] menu.
3. Select
[
Ok
]
directly in the
[
PSV Setup
] menu. Or, you can set [
Backup Mode Active
],
[
Trigger Level
], and
[
PSV
Insp Termination Level
]
followed by selecting [
Ok
].
After
[
Ok
]
is selected, the [
Psupp
] shortcut key (the second key from the right in the
parameter setup shortcut keys area) is highlighted.
4. Make sure that Psupp is appropriately set for the patient. Push the control knob to
confirm the setting so as to start PSV mode.
NOTE
z
When it is necessary to switch over to PSV mode, confirm the setting of Psupp first.
Otherwise, the system works in the previous ventilation mode. If the setting of
Psupp is not confirmed for 10 s, the screen returns to the previous mode
automatically.
z
Before activating a new mechanical ventilation mode, make sure that all related
parameters are set appropriately.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...